Query: What as-found bias is recorded on calibration certificate C-77413? average_relevancy=0.2001 ndcg=1.0000 --- rank 1 by relevancy --- similarity=0.7468 relevancy=0.5140 source=06c5221a-8323-49d4-8bf0-02db17c0c558/4dece290-11ac-46c2-8d2e-99a00b6e2193/A03_calibration_certificate_scan.pdf INSTRUMENT **Tolvan A3 conductivity-temperature sensor** SERIAL NUMBER **FW-A3-40871** CUSTOMER **Severn Estuary Monitoring Partnership** ORDER REFERENCE **SEMP/PO/9134** DATE OF CALIBRATION **3 March 2026** NEXT CALIBRATION DUE **3 January 2028** PROCEDURE **FWL-CAL-114, issue 6** AMBIENT **20.4 C, 61 % RH** --- ## Results Salinity channel, checked against reference standard SR-2 at four points | Reference point | Reference value | As found | As left | | --------------- | --------------- | -------- | ------- | | Point 1 | 5.000 psu | 4.961 psu | 4.997 psu | | Point 2 | 15.000 psu | 14.964 psu | 14.996 psu | | Point 3 | 25.000 psu | 24.962 psu | 24.995 psu | | Point 4 | 35.000 psu | 34.965 psu | 34.997 psu | ## Summary As-found bias, mean of four points: -0.037 psu As-left bias, mean of four points: -0.004 psu Expanded uncertainty (k = 2): 0.008 psu The instrument was adjusted after the as-found check and passed. Conductivity cell and thermistor were cleaned before the as-found run. Signed, calibration engineer M. Ottaway --- rank 2 by relevancy --- similarity=0.6690 relevancy=0.4865 source=06c5221a-8323-49d4-8bf0-02db17c0c558/4dece290-11ac-46c2-8d2e-99a00b6e2193/A10_tolvan_a3_product_overview.md # Tolvan A3 - product overview The Tolvan A3 is a conductivity and temperature sensor built for continuous deployment in estuarine and coastal water. It is the instrument most of our customers meet first, and most of the questions the service desk answers are about this one. ## What it measures The instrument reports conductivity, temperature and a derived practical salinity. Conductivity is measured with an inductive cell, which has no exposed electrodes to foul or to wear, and temperature with a thermistor mounted in the same flow path so that the two readings describe the same parcel of water. Pressure is not measured; where a depth record is needed, the instrument is paired with a separate pressure sensor on the same mooring. ## Where it is used Typical deployments are estuary monitoring for a regulator, intake monitoring for a water treatment works, and baseline surveys ahead of construction in a harbour. It is rated for continuous immersion to fifty metres. Deeper work uses the Tolvan C5, which shares the sensing arrangement in a pressure housing rated considerably deeper. ## How it is powered and read On a mooring, the instrument runs from an internal battery pack and logs to internal memory. On a cabled installation it is powered over the same cable that carries its data, and reports continuously to the Brannoc telemetry gateway, which forwards to the customer's own system. The two modes are configured at commissioning and are not switched in the field. ## Fouling and cleaning In estuarine water the practical limit on deployment is biological fouling rather than power or memory. An anti-fouling guard and a sacrificial anode are fitted at commissioning and replaced at each service visit. Customers in warm or nutrient-rich water are advised to shorten the interval between visits rather than to clean the sensing face themselves, because the inductive cell is easily damaged by an abrasive. ## Calibration Every instrument leaves the laboratory with a signed certificate recording the as-found and as-left bias at four reference points across the working range. The recommended recalibration interval for the Tolvan A3 is 22 months from the date of the last certificate. An instrument returned early because of a suspected drift is checked as found before anything is adjusted, so that the drift can be quantified rather than merely corrected. --- rank 3 by relevancy --- similarity=0.6648 relevancy=0.0000 source=06c5221a-8323-49d4-8bf0-02db17c0c558/4dece290-11ac-46c2-8d2e-99a00b6e2193/A05_depot_and_commissioning_note.md ## Commissioning: before travelling Confirm the site permit is in place and that the structure is accessible at the tide state you have planned for. Confirm the instrument's job number and that the certificate in the case matches the serial number on the housing. Take the torque wrench; the sensing face is the one part of this job that does not tolerate improvisation. ## Commissioning: mounting Mount the instrument on the mounting bracket with the sensing face pointing into the flow, not across it. Leave at least 150 mm of clear water in front of the face. On a structure with a grating, mount below the grating rather than through it, so that the face is not shaded by a bar that will collect weed. Bracket bolts are stainless M10, tightened to 22 Nm. Do not substitute a galvanised bolt: the sacrificial anode is sized for the stainless arrangement and the wrong metal in the water changes what corrodes. ## Commissioning: cable routing Route the deck cable away from any winch path and secure it every 400 mm. Leave a drip loop below the gland. Where the cable passes a sharp edge, fit the sleeve from the gland kit rather than taping the cable. Do not coil surplus cable against the instrument housing; surplus goes in a loose figure of eight on the deck stand. ## Commissioning: checks before leaving site With the instrument in the water and the gateway powered, confirm a reading appears at the gateway within two minutes, and that the temperature reading is within half a degree of a hand-held thermometer in the same water. Confirm the salinity reading is stable over five minutes; a reading that wanders is usually air trapped in the cell, cleared by lowering and raising the instrument twice. Record the gateway signal strength on the commissioning sheet. Leave the certificate in the site file, not in the instrument case. Leave one spare gland kit and one spare anode with the site's own stores, and take the packaging away. Send back the signed commissioning sheet, photographs of the mounting from two angles, and the serial numbers of anything fitted, on the same day. A commissioning sheet that arrives a week later has usually lost the detail that matters. --- rank 4 by relevancy --- similarity=0.6597 relevancy=0.0000 source=06c5221a-8323-49d4-8bf0-02db17c0c558/4dece290-11ac-46c2-8d2e-99a00b6e2193/A04_employee_handbook.docx Employee Handbook Fenwake Instruments Ltd - all locations 1. Working here This handbook sets out the terms that apply to everyone employed by Fenwake Instruments Ltd, whether based at the Bristol workshop, at the Avonmouth calibration laboratory, or working from a customer site or vessel. It forms part of your contract. Where a term is set out in your written statement of particulars and also here, the written statement governs. 2. Hours of work Standard hours are 37.5 a week, worked between 08:00 and 18:00, Monday to Friday. Workshop staff work a 07:30 start by agreement with the workshop manager. Anyone rostered to the out-of-hours duty engineer rota is paid a standby allowance for each week on the rota, and an hourly rate for any hour actually worked while on standby. 2.1 Time recording Hours are recorded against a job number on the day they are worked. A job number is required for chargeable work, for internal projects and for training. Corrections are made by the line manager rather than by editing a submitted week. 3. Annual leave Full-time employees accrue 26 days of annual leave in each complete leave year, which runs from 1 April to 31 March. Part-time employees accrue pro rata. Leave accrues from the first day of employment and may be taken during the probation period with the line manager's agreement. 3.1 Booking leave Annual leave is requested in the rota system and confirmed by the line manager. Requests are considered in the order they are received. Service operations will decline a request that would leave the duty engineer rota uncovered, and will say so within five working days so that alternative dates can be chosen. 3.2 Carry-over Up to five days of untaken annual leave may be carried into the next leave year, and must be taken by 30 June. Anything beyond five days is carried over only where leave could not be taken because of a business requirement, and only with the written agreement of a director. 3.3 Leave on leaving On termination, accrued and untaken annual leave is paid in the final salary payment. Where more leave has been taken than accrued, the balance is deducted from that payment. 4. Sickness absence --- rank 5 by relevancy --- similarity=0.6618 relevancy=0.0000 source=06c5221a-8323-49d4-8bf0-02db17c0c558/4dece290-11ac-46c2-8d2e-99a00b6e2193/A08_warranty_terms_sales_pack.pdf Sales pack Page 1 Fenwake Instruments Ltd Warranty Terms Sales pack - customer-facing warranty terms 1. What is covered These are the warranty terms that accompany a new instrument. They apply to every new Tolvan A3 supplied by Fenwake Instruments Ltd, however it was ordered and wherever it is deployed. The warranty covers the instrument, its pressure housing, its connector and any cable supplied with it. It covers failure arising from a defect in materials or workmanship. Where a covered failure occurs, Fenwake repairs or exchanges the instrument at its own option and bears the cost of carriage in both directions. 2. The warranty period Every new Tolvan A3 carries a warranty of 24 months from the date the instrument is delivered. The period runs from delivery rather than from commissioning, and is not extended by a repair carried out under it. 2.1 Instruments repaired under warranty A repaired instrument carries the unexpired balance of the original period, or ninety days from the return of the repaired instrument, whichever ends later. 3. What is not covered • Fouling, corrosion of sacrificial components, and wear of consumables such as anodes, guards and gland kits. • Damage caused by a mooring failure, a vessel strike, a snagged cable or any other mechanical event outside the instrument. • Damage caused by opening the housing outside a Fenwake workshop, or by breaking a calibration seal. • Drift within the published specification. Drift is a calibration matter, not a defect. • Any instrument whose serial number has been removed or altered. 4. Making a claim A claim is made through the service desk, quoting the serial number and the delivery note. The service desk raises a ticket, arranges return carriage and tells the customer within five working days whether the failure is covered. Where it is not covered, the customer is quoted for the repair before any work starts. 4.1 If the customer disagrees Where a customer disputes a decision that a failure is not covered, the instrument is held unrepaired and the head of service operations reviews the file. The instrument is not returned, scrapped or repaired while a review is open. Fenwake Instruments Ltd - Warranty Terms Sales pack Page 2