# Tolvan A3 - product overview

The Tolvan A3 is a conductivity and temperature sensor built for
continuous deployment in estuarine and coastal water. It is the instrument most
of our customers meet first, and most of the questions the service desk answers
are about this one.

## What it measures

The instrument reports conductivity, temperature and a derived practical
salinity. Conductivity is measured with an inductive cell, which has no exposed
electrodes to foul or to wear, and temperature with a thermistor mounted in the
same flow path so that the two readings describe the same parcel of water.
Pressure is not measured; where a depth record is needed, the instrument is
paired with a separate pressure sensor on the same mooring.

## Where it is used

Typical deployments are estuary monitoring for a regulator, intake monitoring
for a water treatment works, and baseline surveys ahead of construction in a
harbour. It is rated for continuous immersion to fifty metres. Deeper work uses
the Tolvan C5, which shares the sensing arrangement in a
pressure housing rated considerably deeper.

## How it is powered and read

On a mooring, the instrument runs from an internal battery pack and logs to
internal memory. On a cabled installation it is powered over the same cable
that carries its data, and reports continuously to the Brannoc
telemetry gateway, which forwards to the customer's own system. The two modes
are configured at commissioning and are not switched in the field.

## Fouling and cleaning

In estuarine water the practical limit on deployment is biological fouling
rather than power or memory. An anti-fouling guard and a sacrificial anode are
fitted at commissioning and replaced at each service visit. Customers in warm
or nutrient-rich water are advised to shorten the interval between visits
rather than to clean the sensing face themselves, because the inductive cell is
easily damaged by an abrasive.

## Calibration

Every instrument leaves the laboratory with a signed certificate recording the
as-found and as-left bias at four reference points across the working range.
The recommended recalibration interval for the Tolvan A3 is 22 months from the date of the last certificate. An instrument returned early because of a
suspected drift is checked as found before anything is adjusted, so that the
drift can be quantified rather than merely corrected.

## Firmware

Firmware is updated during a service visit rather than over the telemetry link,
because an interrupted update on a deployed instrument leaves it unreadable
until it is recovered. A customer who declines a recommended update is asked to
confirm in writing, and the confirmation is attached to the instrument record.